Hello, welcome to visit the official website of Shanghai Jifa Instrument Equipment Co., Ltd.!

Fabric Performance Tester

DHTH-100-40-P-SD High and Low Temperature Alternating Humidity and Heat Test Box

product manual

1 Key technical indicators Main Technical IndicatorsTest space volume Available test space volume100l about100riseTest space dimensions available

  • description

  • compliant

  • Parameters

  • principle

  • Standard


Key technical indicators

Test space volume 

About 100 litres.

Test space dimensions 

D500x W400x H500mm (Deep x wide x high)

Temperature range

-40 to +100ºC

 

Competitive Edges and Standard Features 

Substitute for imported chambers from Germany, American and Japan

Quarter of those chambers 

Same quality with those chambers 

Same test parameters with those chambers 

Specimen protection against condensation and over-tempering 

Moveable and height adjustable design 

2 standard entry ports (Ø50 mm) 

2 insert shelves 

Fault diagnosis system for maintenance and repair

Observation window in door incl. test space lighting 

Air-cooled refrigeration system 

Networking capabilities via RS 232 or TCP/IP 

Powder-coated housing and stainless steel inner test space

Intelligent airflow for high accuracies at various loadings 

Humidity Chambers (TH-Models) only features: 

Proprietary psychrometric measurement system 

Humidity water tank with water level indicator 

Automatic replenishment feature(optional) 

Special port Plugs For 85/85 tests 

 


 

Test Parameters 

3.1 Temperature


Temperature range 

-40 to +100ºC

Temperature deviation in time in center of working space acc. to IEC 60068-3-5

±0.1 to ±0.5 K (in time) 

Temperature homogeneity in space relative to the set value acc. to IEC 60068-3-5

±0.5 to ±2.0 K (in space) Evenness

Temperature rate of change acc.to IEC 60068-3-5

2.0--3.0 K/min (heating) Warming up (no-load, non-linear)

0.7--1.0K/min (cooling) Cooling (no-load, non-linear)

 

 

3.2 Humidity 

Temperature range 

+15 to +95ºC

Humidity range 

20%to 98% r.h

Humidity accuracy 

±1.0 to ±3.0% r.h.

Water consumption 

< 2 l/24 h at +40ºC, 92% r.h.



HHumidity range 

Humidity diagram (A: temperature, B: relative humidity)


 

Chamber Design

Test Space 

 

500 mm deep  

400 mm wide  

500 mm high  

 

Test Space Dimensions 


 

Material 


stainless steel grade sus304

Chamber loading 


max. 30 kg per shelf 

max. 80 kg per shelf 

max. 50 kg on floor 

max. 130kg total

Shelf 

 

 

 


2 pieces made of stainless steel,

max. 8 numbers

Entry ports



Ø50 mm 2 ports with plugs

(2 ports in total) 



 

Window (heated)


300mm wide 

400mm high 

Door


hinged to the left, single hand operation        

 

 System and Installation 

Cabinet dimensions (incl. touch panel)


1285mm deep 

600mm  wide 

1580mm high  

Weight 


310kg (net) 

Sound level 


60 dB(A) (1m in front of chamber

Ambient operation conditions


+10 to +35ºC

max. 75% r.h.

Chamber housing 


stainless steel inner test space

Entry port locations


Standard: R1 Ø50 mm,L1 Ø50 mm

Optional

R2…Rn locations right-hand side

L2…Ln locations left-hand side

D1 location in chamber ceiling

 

Supplies & Connections 


Nominal voltage 

3/PE AC 380/400 V ±10%, 50 Hz

Nominal power 

4.0KW

Mains connection 

3M power supply 6mm 2x5 core

Drainage 

¾" hose connections, back-pressure free for condensation water

Humidification water 

city water

Refrigeration system 

air-cooled mechanical refrigeration system based on

Freon R 404 A with a ozone depletion potential of zero (ODP = 0.00)

Control and Programming


Programming

Program memory

max. 50 programs 

PID controller

PID 

special control algorithms defined according to test

tasks 

Segments 

max. 99 

Loops 

10 (nested) 

Program cycles

max. 99 

Password protection

two levels, to prevent accidental settings 

Limit value monitoring system


for temperature

Diagnostic system

for information on operating times and possible operating failures

 

 

USB Interface Export Data Function

for connection to a host computer system

(e.g. Notebook-operating panel or

VNC VIEWER)

 

  Operation with Color Touch Panel 

  The color touch panel, suitable for graphics with a resolution of 800 x 480 Pixel (VGA), is part 

  of the standard equipment. It allows a convenient input of fixed values and program operation 

  with graphic display of set points and actual values, the operating time and number of remaining 

  cycles etc. Furthermore, the touch panel has the following special features:

 

  Background-lit display

  Menu-guidance

  Easy programming of individual test programs

  Safe storing of individual programs, which can be activated at any time

  

  Operation state displayed by means of light diodes

  Languages (switchable) 

  Standardized display of all parameters (temperature, humidity etc.)

  

  Technical Details

Construction

The climate test chambers are supplied ready for connection. Compact design means a minimum of space is required. Three test chambers with a volume from 100 to 1000 l are available. All chambers are designed to be free-standing in one location.

The machine section incorporating the refrigeration unit, the water storage tank and the water pump is located below the test chamber. The swing-out switch cabinet is located on the rear of the cabinet. The operating elements and the main switch are located on the right side of the cabinet.


Exterior Housing

The exterior housing of the unit is made from lightweight corrosion-resistant, self-supporting, galvanized sheet-steel.

The environmental-friendly insulation between test chamber and exterior housing guarantees best insulation values and thus lowest possible operating costs.

 

Test Chamber

The test chamber consists of a stainless-steel container and is welded vapor-tight. The test chamber is easy to clean thanks to special welding, smooth surfaces, rounded edges and shelf supports. The rear test chamber wall incorporates an air duct with built-in humidifier tub (C models), built-in heat exchanger, electric heaters and a recirculating air fan. Air guidance via baffle plates in the test chamber floor and ceiling produces optimum ambient air and temperature distribution.

 

Test Chamber Door

The test chamber is completely sealed by a door that is hinged on the left. It is provided with an observation window (not for ECO version). The door opens fully for easy access and is equipped with high quality insulation. An optimum seal is guaranteed by special profile seals and a self-tightening locking mechanism which is operable with one hand.

 

Access Ports

All chambers have two access ports (approx. 50mm diameter) made of low thermal conductivity material. The ports are in the left and right side walls and can be used for measuring and control wires, other supply connections or accessories.

If not required, these ports are closed with rubber plugs supplied as standard.

 

Temperature and Humidity Conditioning

The large axial fan with exterior drive motor draws the air out of the space. This recirculating air then passes a fin-type heat exchanger, which cools the air, if necessary. The special design of the refrigerating system avoids unintended condensation precipitation at the heat exchanger during climatic operation and guarantees highest temperature and humidity constancies in time. An electrical heater, installed in front of the heat exchanger in air direction, is heating-up the recirculating air.

An innovative humidification system allows a high relative humidity even with a heat load (C models).

 

Protection System against Condensation

The protection system consists of a dehumidifying evaporator which prevents condensation forming on the specimens as they heat-up after tests at low temperatures.

The protection system's operating range is -40 to +60 °C.

The protection system against condensation is activated by a digital switch channel. It is possible to switch on the protection system manually or to integrate it into a program.

 

Permanently Memorized Overheating Safety Thermostat

A steady state overheating safety thermostat is installed that automatically switches off the heater if the test chamber's maximum temperature range is exceeded.

 

Power Supply

Each functional circuit is equipped with its own safety device, which, in the event of trouble, turns off the affected circuit or the entire cabinet. The nature of the trouble is visually displayed.

Wiring and electrics are governed by the latest technology and strictly conform to safety regulations for electrical installations and materials, as well as to relevant VDE regulations.

The test chamber are standard-wise equipped with a lockable main switch according to VDE.

 

  General Statements 

In the constant effort to improve our product, we reserve the right to make construction or design changes without prior notice or obligation.

 

Please note that for installation of the climate test cabinet (for components which jut out, such

as: hinges, locks, etc.) the clearance requirements indicated in the technical data must be observed.

 

The performance data indicated were determined with empty test space and refer to an appliance without additional ports or devices, which represent a thermal load with an ambient temperature of max. +25 °C and with water-cooled refrigeration units (option) to a water temperature of approx. +28 °C.

 

The admissible ambient temperature for operation of the cabinet is between +10 °C and +35 °C (+50 °F and +95 °F).

 

Performance data cannot be guaranteed in ambient conditions other than those stated.

 

Accuracy rates are proven in the test space center, in steady state, without specimens, without radiation and without additional equipment in the test space. The cooling down rate stated refers to the temperature range without controlled humidity.

 

Noise measurements and sound level data are effected as per DIN 45 635, part 1, accuracy class 2.

 

In case of outgassing or pollution caused by the specimen we recommend regular maintenance of the measuring system. Please keep in mind that outgassing of specimens and connection lines might cause corrosion of components. Corrosive gasses might escape among others from chip boards, cardboard boxes and bandaging materials. If outgassing is to be expected, we recommend an optional ventilation of the test space with a purge air volume. If outgassing or pollution is expected, we also recommend a regular maintenance of the measuring system。

 

Measures to be taken regarding a necessary combination with other units have to be approved by the local supervising authorities. The relevant costs are not included in our quotation.

 

Please note that the sealing material and cable insulation used in our test units and systems contains silicone. If this is a problem for you, please contact us.

The EMC test and the data on the noise emission are in accordance with EN 50 081-1 and EN61 000-6-2.


Shanghai Jifa Instrument Equipment Co., Ltd. Official WeChat
Add:

2/F, Building 3, No. 3599, Yuanjiang Road, Minhang District, Shanghai

Factory address:

No. 1, Lane 19, Zhongyi East Road, Yucheng Town, Jinshan District

Fax:

021-33321965

E-mail:

jifayiqi@163.com

Copyright © 2019 Shanghai Jifa Instrument Equipment Co., Ltd. Technical support: Jiedian Technology