TSI8130 automatic filter material detector can measure the filtration efficiency of filter material and mask quickly and accurately. Model 8130 produces sodium salts or oily aerosols as test carriers, which can test filter materials and masks with 99.999% filtration efficiency, and conform to the specifications of NIOSH Regulation 42 CFR (part 84). During the outbreak of SARS in 2003, the technical specifications of 8130 were officially listed as the national standard GB 19083-2003 for technical requirements of medical protective masks.
1. American Commercial Respiratory Regulations (CFR 42, Part 84)
2. New Functions of EU EN 143 and Respiratory Standards
3. GB 19083-2003 Technical Requirements for Medical Protective Masks
4. GB 2626-2006 "Respiratory protective articles self-priming filter type anti-particulate respirator"
5. JMOL (Japanese Respiratory Standard) and Multinational Respiratory Regulations, etc.
1. Aerosol generation:
2. Oiliness (8127 and 8130):
3. Aerosol types: PAO, DOP, DEHS, paraffin
4. Median mass diameter: 0.33 um
5. Quantitative median diameter: 0.20 um
6. Geometric standard deviation: <1.6
7. Aerosol Concentration: 50-200 mg/m3
8. Salt type (8130 only):
9. Aerosol Type: Sodium Chloride
10. Median mass diameter: 0.26 um
11. Quantitative Median Diameter: 0.075 um
12. Geometric standard deviation: <1.83
13. Aerosol Concentration: 12-20 mg/m3
14. Aerosol detection:
15. Detection Method: Dual Solid State Photometer
16. Concentration measurement range: 1.0 ug/m3-200 mg/m3
17. Flow range: 15-100 L/min
18. Measurement range of efficiency: 0-99.999%
19. Requirements for Compressed Gas Source: Flow Rate 198 L/min at 550 kPa
2/F, Building 3, No. 3599, Yuanjiang Road, Minhang District, Shanghai
No. 1, Lane 19, Zhongyi East Road, Yucheng Town, Jinshan District
021-33321965
jifayiqi@163.com