Hello, welcome to visit the official website of Shanghai Jifa Instrument Equipment Co., Ltd.!

Imported Instrument Supplies

Q-Sun B02 Solar Color Fastness Instrument

product manual

The design of Q-Su n B02 sunshine color fastness tester is based on the requirements of ISO 105 B02 and ATCC method 16 sunshine color fastness test

  • description

  • compliant

  • Parameters

  • principle

  • Standard

The design of Q-Su n B02 sunshine color fastness tester is based on the requirements of ISO 105 B02 and ATCC method 16 sunshine color fastness test. It has a drum-type sample holder, which provides accurate control of key parameters such as spectrum, irradiance, relative humidity, temperature in the box and blackboard temperature. Compared with similar equipment, the test capacity is 48% to 92%.

AATCC TM16, ISO 105 B02, GB8427

1. Blackboard temperature

Light cycle - B/SL filter 50 oC - 115 oC

Light cycle - IR filter 35 oC - 100 oC

Dark Cycle 25 oC - 45 oC

2. Black Standard Temperature

Light cycle - B/SL filter 55 oC - 125 oC

Light cycle - IR filter 40oC - 110 oC

Dark Cycle 25 oC - 45 oC

3. Male illumination

IR filter 1.10-0.20 W/m2@ 420 nm

42-10 W/m2@ TUV 300-400 NM

B/SL filter 1.10-0.20 W/m2@ 420 nm

48-10 W/m2@ TUV 300-400 NM

4. Relative humidity

Maximum relative humidity 60% - 20%.

(When the blackboard temperature is below 60 oC)

5. External dimensions

Height 66.0 inches 168 cm

Width 36.5 inches 93 centimeters

Depth 27.0 inches 69 centimeters

6, weight

Net weight 328 pounds 149 kilograms

Gross weight 520 pounds 236 kilograms

7. Size of test room

Height 15.8 inches 40 centimeters

Width 12.5 inches 32 centimeters

Deep 16.8 inches 43 centimeters

8. Sample area

Height 11.9 inches 30 centimeters

9.8 inches in diameter, 25 centimeters

Perimeter 30.8 inches 78 centimeters

Total area 367 square inches 2367 square centimeters

9. Sample rack size

Size 1.8 x 10.4 inches 46 x 264 mm

Area 18.7 square inches 121 square centimeters

Total exposure area 290 square inches 1882 square centimeters

10. Requirements for Power Supply and Water Supply

208 or 230 V single-phase power supply, 50 - 60 Hz, 25 Amperes

Deionized water, consumption 1 liter/hour


1. Xenon lamp tube

B02 uses an air-cooled xenon lamp tube to provide full spectrum sunlight with a power of 1800 W. Q-Lab recommends replacing the lamp every 1500 hours.

2. Optical filter

The optical filter consists of borosilicate glass outer filter and 14 inner filter. The inner filter is arranged in two groups of heptagons. According to the requirements of exposure or test conditions, choose the inner filter, Windows-IR or 3, Windows-B/SL.

3. Irradiation Control

B02 is equipped with solar eye irradiance control system, which can refer to the setting range of irradiance in technical parameters.

Temperature and relative humidity control

B02 can control the temperature of the box and blackboard or black mark at the same time. The humidity sensor can accurately control the relative humidity and refer to the setting range of relative humidity in the technical parameters.

4. Calibration

Auto Cal: Calibration using CR20 radiometer.

Blackboard temperature: Calibration with CT202 thermometer.

Box temperature and relative humidity: sensor is replaced once a year.

All calibrations can be traced back to NIST and meet ISO requirements.

CR20 and CT202 need to be sent back to Q-Lab company for calibration every year.


Q-SUN B02主机(420nm校准点、基本型)
Includes 包含:
1、Solar Eye Irradiance Control.太阳眼光辐照度控制系统
2、Chamber Air Temperature Control.箱体温度控制系统
3、Precise relative humidity control.精确箱体湿度控制系统
4、Standard xenon lamp (1800w×1500hr)标准氙灯(1800w×1500小时)
5、Specimen holders Kits.大容量样品架(可容纳31个样品夹,155块测试样品)
6、CR20 Calibration Radiometer (420nm)光辐照度传感器标定器(CR20)
ISO 105 B02 Kits(ISO 105 B02标准操作包)
1、Lantern with Window-IR filters.Window-IR内外滤光片套筒
2、Black Standard Temperature Sensor.黑标温度传感器BST
3、Textile Metal Masks for ISO (15 long,1 short each for both 1/3 mask and 2/3 mask)ISO样品金属遮盖板一套
AATCC TM16-3 Kits(AATCC TM16-3标准操作包)
1、Lantern with Window Glass-B/SL filter for AATCC TM16-3.Window-B/SL内外滤光片套筒(符合AATCC 16-3标准)
2、Black Panel Temperature Sensor.黑板温度传感器(符合AATCC 16-3标准)
3、Textile Masks for AATCC (Including pack of 100 masks, 31 pieces window frames, and 15 center conversion screws)AATCC样品遮盖板一套(包含纸卡一包100片,31个压框,15个转换螺丝)

Shanghai Jifa Instrument Equipment Co., Ltd. Official WeChat
Add:

2/F, Building 3, No. 3599, Yuanjiang Road, Minhang District, Shanghai

Factory address:

No. 1, Lane 19, Zhongyi East Road, Yucheng Town, Jinshan District

Fax:

021-33321965

E-mail:

jifayiqi@163.com

Copyright © 2019 Shanghai Jifa Instrument Equipment Co., Ltd. Technical support: Jiedian Technology