The design of Q-Su n B02 sunshine color fastness tester is based on the requirements of ISO 105 B02 and ATCC method 16 sunshine color fastness test. It has a drum-type sample holder, which provides accurate control of key parameters such as spectrum, irradiance, relative humidity, temperature in the box and blackboard temperature. Compared with similar equipment, the test capacity is 48% to 92%.
AATCC TM16, ISO 105 B02, GB8427
1. Blackboard temperature
Light cycle - B/SL filter 50 oC - 115 oC
Light cycle - IR filter 35 oC - 100 oC
Dark Cycle 25 oC - 45 oC
2. Black Standard Temperature
Light cycle - B/SL filter 55 oC - 125 oC
Light cycle - IR filter 40oC - 110 oC
Dark Cycle 25 oC - 45 oC
3. Male illumination
IR filter 1.10-0.20 W/m2@ 420 nm
42-10 W/m2@ TUV 300-400 NM
B/SL filter 1.10-0.20 W/m2@ 420 nm
48-10 W/m2@ TUV 300-400 NM
4. Relative humidity
Maximum relative humidity 60% - 20%.
(When the blackboard temperature is below 60 oC)
5. External dimensions
Height 66.0 inches 168 cm
Width 36.5 inches 93 centimeters
Depth 27.0 inches 69 centimeters
6, weight
Net weight 328 pounds 149 kilograms
Gross weight 520 pounds 236 kilograms
7. Size of test room
Height 15.8 inches 40 centimeters
Width 12.5 inches 32 centimeters
Deep 16.8 inches 43 centimeters
8. Sample area
Height 11.9 inches 30 centimeters
9.8 inches in diameter, 25 centimeters
Perimeter 30.8 inches 78 centimeters
Total area 367 square inches 2367 square centimeters
9. Sample rack size
Size 1.8 x 10.4 inches 46 x 264 mm
Area 18.7 square inches 121 square centimeters
Total exposure area 290 square inches 1882 square centimeters
10. Requirements for Power Supply and Water Supply
208 or 230 V single-phase power supply, 50 - 60 Hz, 25 Amperes
Deionized water, consumption 1 liter/hour
1. Xenon lamp tube
B02 uses an air-cooled xenon lamp tube to provide full spectrum sunlight with a power of 1800 W. Q-Lab recommends replacing the lamp every 1500 hours.
2. Optical filter
The optical filter consists of borosilicate glass outer filter and 14 inner filter. The inner filter is arranged in two groups of heptagons. According to the requirements of exposure or test conditions, choose the inner filter, Windows-IR or 3, Windows-B/SL.
3. Irradiation Control
B02 is equipped with solar eye irradiance control system, which can refer to the setting range of irradiance in technical parameters.
Temperature and relative humidity control
B02 can control the temperature of the box and blackboard or black mark at the same time. The humidity sensor can accurately control the relative humidity and refer to the setting range of relative humidity in the technical parameters.
4. Calibration
Auto Cal: Calibration using CR20 radiometer.
Blackboard temperature: Calibration with CT202 thermometer.
Box temperature and relative humidity: sensor is replaced once a year.
All calibrations can be traced back to NIST and meet ISO requirements.
CR20 and CT202 need to be sent back to Q-Lab company for calibration every year.
Q-SUN B02主机(420nm校准点、基本型)
Includes 包含:
1、Solar Eye Irradiance Control.太阳眼光辐照度控制系统
2、Chamber Air Temperature Control.箱体温度控制系统
3、Precise relative humidity control.精确箱体湿度控制系统
4、Standard xenon lamp (1800w×1500hr)标准氙灯(1800w×1500小时)
5、Specimen holders Kits.大容量样品架(可容纳31个样品夹,155块测试样品)
6、CR20 Calibration Radiometer (420nm)光辐照度传感器标定器(CR20)
ISO 105 B02 Kits(ISO 105 B02标准操作包)
1、Lantern with Window-IR filters.Window-IR内外滤光片套筒
2、Black Standard Temperature Sensor.黑标温度传感器BST
3、Textile Metal Masks for ISO (15 long,1 short each for both 1/3 mask and 2/3 mask)ISO样品金属遮盖板一套
AATCC TM16-3 Kits(AATCC TM16-3标准操作包)
1、Lantern with Window Glass-B/SL filter for AATCC TM16-3.Window-B/SL内外滤光片套筒(符合AATCC 16-3标准)
2、Black Panel Temperature Sensor.黑板温度传感器(符合AATCC 16-3标准)
3、Textile Masks for AATCC (Including pack of 100 masks, 31 pieces window frames, and 15 center conversion screws)AATCC样品遮盖板一套(包含纸卡一包100片,31个压框,15个转换螺丝)